P845 Long Shot Width
Adjust the width (in ms) of the long shot transmit pulse.
Primary Index
Values
Altered By
Related
This feature is automatically altered when Transducer (P004) is altered.
P850 Short Shot Bias
Use this feature to slant the echo evaluation in favour of the short shot echo when both
short and long shots are evaluated (see Shot Mode, P803).
Primary Index
Values
Related
P851 Short Shot Floor
Enter the minimum echo strength (in dB above 1 uV) derived from a short shot to be
considered for evaluation.
Primary Index
Values
Related
Page 214
MultiRanger 100/200 – INSTRUCTION MANUAL
Single Point Model
Global
Range: 0.000 to 5.000
•
P004 Transducer
•
P841 Long Shot Number
•
P844 Short Shot Width
•
P843 Long Shot Frequency
Single Point Model
Global
Range: 0 to 100
Preset: 20
•
P803 Shot / Pulse Mode
•
P840 Short Shot Number
•
P842 Short Shot Frequency
•
P844 Short Shot Width
•
P851 Short Shot Floor
•
P852 Short Shot Range
Single Point Model
Global
Range: 30 to 100
Preset: 50
•
P840 Short Shot Number
•
P842 Short Shot Frequency
•
P844 Short Shot Width
•
P850 Short Shot Bias
•
P852 Short Shot Range
Dual Point Model
Transducer
Dual Point Model
Transducer
Dual Point Model
Transducer
7ML19981FB05