8. Commissioning
Limit switches are available for both digital and relay outputs.
Direction mode: 1 set point at 0% flow; hysteresis 5%.
If 2 set points must activate 2 separate outputs, a single set point has to be selected individually for
digital as well as relay outputs.
Note
When batch function is on relay - the pulse/frequency output is not possible.
The relay output must always be used to operate the cleaning unit when a cleaning unit has been
installed together with the transmitter. The relay output cannot be used for other purposes.
Batch control is available on MAG 6000 only.