Reading Measurements by Operation
P001
0 – OFF
1 – Level
2 – Space
3 – Distance
4 – Difference [MR 200]
5 – Average [MR 200]
6 – OCM [MR 200]
7 – Pump Totalizer [MR 200]
P921 Material Measurement
The distance in Units (P005) or % of Span (P007) between Empty (P006) and the
monitored surface.
Primary Index
Values
Related
P922 Space Measurement
The distance between the monitored surface and Span (P007).
Primary Index
Values
Related
P923 Distance Measurement
The distance between the monitored surface and the transducer face.
Primary Index
Values
7ML19981FB05
MultiRanger 100/200 – INSTRUCTION MANUAL
P050 = 0
- - - -
P921
P922
P927
P921 (indexed)
P921 (indexed)
P925
P924
Level
Range: -999 to 9999
P005 Units
P006 Empty
P007 Span
Transducer
Range: 0.000 to 9999
P007 Span
Transducer
Range: 0.000 to 9999
P050 ≠ 0
- - - -
P924
100% - P924
P927
P921 (indexed)
P921 (indexed)
P925
P924
Page 219