Siemens MAG 1100 Operating Instructions Manual page 181

Electromagnetic flowmeters
Table of Contents

Advertisement

Cable
• Do not use coaxial cable for transducer (see General Appendix F: Upgrading on
page 246 for more information)
• transducer and mA output signal to be 2 copper conductors, twisted with shield/drain
wire, 300 Vrms, 0.5 mm
conductors @ 1kHz = 62.3 pF/m (19 pF/ft). Nominal capacitance between conductor
and shield @ 1kHz = 108.3 pF/m (33 pF/ft) (Belden®
• 365 m maximum
The MultiRanger is to be used only in the manner outlined in this instruction
Note:
manual or protection provided by the equipment may be impaired.
1.
Belden is a registered trademark of Belden Wire & Cable Company.
7ML19981FB05
MultiRanger 100/200 – INSTRUCTION MANUAL
2
(22-18AWG), nominal capacitance between adjacent
1
8760 is acceptable)
Page 7