Siemens MAG 1100 Operating Instructions Manual page 326

Electromagnetic flowmeters
Table of Contents

Advertisement

P251 0 or 4 mA Input Level [MR 200]
Shows the process level corresponding to the 0 or 4 mA value.
Primary Index
Values
Related
When using an external mA signal to determine level, the input range must be scaled to
give accurate results.
P252 20 mA Input Level [MR 200]
Shows the process level corresponding to the 20 mA value.
Primary Index
Values
Related
Input range is scaled for accuracy if an external mA signal calculates level.
P253 Input Filter Time Constant [MR 200]
Shows the time constant used in the mA input filter to dampen signal fluctuations.
Primary Index
Values
This number in seconds is used in the damping calculations. Larger values damp more
than smaller values and 0 disables the signal filter.
P254 Scaled mA Input Value [MR 200]
Shows the resulting level value after scaling.
Primary Index
Values
This parameter is calculated from the input mA signal.
Page 152
MultiRanger 100/200 – INSTRUCTION MANUAL
Global
Range: -999 to 9999%
Preset: 0%
P006 Empty
P007 Span
Global
Range: -999 to 9999%
Preset: 100%
P006 Empty
P007 Span
Global
Range: 0 to 9999
Preset: 1
Global
Range: -999 to 9999% (view only)
Preset: calculated from the input mA signal
7ML19981FB05