P260 mA Raw Input [MR 200]
Shows the raw mA input supplied by the external device.
Primary Index
Values

Discrete Input Functions (P270 to P275)

Discrete inputs can be used for the following:
Passing other information to a remote system through communications
Backup level override
Use the parameters listed above to have discrete inputs modify the unit's operation.
Use the following parameters to configure the discrete input itself.
Pump Control
See also the
pump control algorithms, including how the discrete inputs alter its operation.
P270 Discrete Input Function
Sets how discrete signals are interpreted by the MultiRanger.
Primary Index
Values
Related
7ML19981FB05
MultiRanger 100/200 – INSTRUCTION MANUAL
mA input
Range: 0.000 to 20.00 (view only)
section on page 57 for a description of the MultiRanger's
Discrete input
0
Forced OFF
1
Forced ON
2
*
Normally Open – 0 (DI open), 1 (DI closed)
3
Normally Closed – 0 (DI closed), 1 (DI open)
Pump Control section
Page 153