Print Is Faint - Thermo Scientific PrintMate A79510100 Operator's Manual

Table of Contents

Advertisement

PrintMate A79510100 Issue 2

Print is Faint

Error Message / Symptom
Print is faint
60
Possible Cause
Cassette Clamp error.
Tape jammed.
Incompatible or old Tape.
Incompatible Cassette.
Instrument Malfunction.
Remedy
Open Print Door and print
a Cassette.
Check Cassette is held
securely during printing and
does not move until the
Cassette Clamp releases it.
If movement is seen call the
Thermo Fisher Scientific
Service Department.
Open Print Door and Waste
Spool Access.
Use Tape Feed function on
Touch Panel to check that
Tape feeds correctly.
If Tape does not feed
remove and refit Tape (see
Chapter 2 - Changing the
Foil Tape).
Replace Tape with approved spare
(see Appendix A - Spares and
Accessories).
Check that Cassette type is included
on Appendix C - Approved Cassette
List.
Call the Thermo Fisher Scientific
Service Department.