Dispense Lever Fails To Reset - Thermo Scientific PrintMate A79510100 Operator's Manual

Table of Contents

Advertisement

PrintMate A79510100 Issue 2

Dispense Lever Fails to Reset

Error Message / Symptom
Dispense lever fails to reset
48
Possible Cause
Cassette blockage when exiting
Hopper.
Cassette blockage in Hopper.
Cassette not compatible.
Other Hopper Malfunction
Instrument Malfunction
Remedy
Open the Print Door and
remove any visible Cassette.
Use the Touch Panel or PC
to continue.
Remove the Hopper at the
output position and use the
Touch Panel or PC to
perform a Dispense Cycle
and reset the dispense lever.
Remove the Cassette Tube
and / or any loose Cassettes
in the Hopper.
Manually open and close
the Hopper Slider a few
time to ensure it moves
freely.
Carefully reload the
Cassettes and refit the
Hopper to the Carousel.
Use the Touch Panel or PC
to continue.
If using a non-approved Cassette,
check that they manually dispense
properly.
Fit another Hopper in the
output position and check
that it operates correctly.
If it does, this indicates that
either the Cassettes or the
Hopper is faulty.
Call the Thermo Fisher Scientific
Service Department.