Tape Errors

Error Message / Symptom
Tape errors
Possible Cause
Tape feed error
Operator Guide
Remedy
Check whether Tape has
run out and replace if
necessary (see Chapter 2 -
Changing the Foil Tape).
Check whether Tape is
slack - use the Tape Feed
function on the Touch
Panel to remove any slack if
present.
Check Tape is fed through
instrument correctly (see
Chapter 2 - Changing the
Foil Tape).
Check Waste Spool rotates
after each Cassette is
printed - If Waste Spool
does not rotate call the
Thermo Fisher Scientific
Service Department.
Cut Tape and remove used
section from the Waste
Spool. Refeed remaining
unused Tape as per
Chapter 2 - Changing the
Foil Tape.