System Specification - Thermo Scientific PrintMate A79510100 Operator's Manual

Table of Contents

Advertisement

PrintMate A79510100 Issue 2

System Specification

Mechanical Specification
Note
Parameters listed are for the 12 hopper model.
810 mm (with cassette tubes) / 290 mm
Height
(without cassette tubes)
Width
435 mm
Depth
380 mm
Weight
17 kg (with cassette tubes)
Electrical Specification
Voltage
100-120 V~ /
220-240 V~
Frequency
50 / 60 Hz
Power
150VA (512
BTU/h)
Earth
< 500 A
Leakage
Fuses
T2A H250V (2
off)
Environmental Specification
Temperature (Operating
Limits)
Temperature (Recommended
Operation)
Temperature (Storage)
Relative Humidity
Altitude
Pollution Degree
Over Voltage Category
10
Warning - For indoor use only
+5°C to +40°C (+41°F to +104°F)
+15°C to +30°C (+59°F to +86°F)
Note Performance may deteriorate when operated
-25°C to +55°C (-13°F to +131°F) / +70°C (+158°F)
Max. 80% RH up to 31°C, decreasing linearly to
Up to 2,000 m (6,500 ft)
outside of this range
for short exposure
50% RH at 40°C
2
II

Advertisement

Table of Contents
loading

Table of Contents