During the working hours on the site, if an immediate communication of the recorded
•
values is needed
Set-up and activate the SMS feature (for the Alarm notification), and test it with the
•
dedicated button.
Check/insert the data network settings.
•
For more information, please refer to chapter 4.3.5.
STEP 3. Acquisition parameters
Choose the desired filter for the velocity sensor.
•
Choose the sampling rate. We suggest to select a sampling rate which does not exceed
•
10 times the maximum frequencies expected in the measurement.
For further details, please refer to chapter 4.4.2.
STEP 4. Trigger recording
Choose the pre- and post- event duration.
•
Choose the max length.
•
Choose the trigger combination. We suggest to trigger when the threshold on one of the
•
three axes is exceeded (X or Y or Z)
We suggest not to activate the trigger increment
•
We suggest to choose the trigger levels after a site evaluation.
•
For further details, please refer to chapter 4.1.2.
MR3000CMR3000C_TR_v2.0_rev1.odt
Figure 5.1.3. Settings of acquisition parameters.
subject to technical change without notice