13.3

Replacing the diaphragm

Third-party spare parts for the pumps can lead to prob‐
lems when pumping.
Only use original spare parts.
Use the correct spare parts kits. In the event of
doubt, refer to the exploded views and ordering
information in the appendix.
Personnel:
n
Requirements:
Put in place protective measures, if necessary.
n
Adhere to the material safety data sheet for the feed chemical.
n
Ensure that the system is at atmospheric pressure.
n
1.
Drain the liquid end: turn the liquid end upside down and allow the
feed chemical to drain; use a suitable medium to thoroughly flush
the liquid end.
2.
With the pump running, move the stroke adjustment dial to the stop
at 0% stroke length.
ð The drive axle is now difficult to turn.
3.
Switch off the pump.
4.
Unscrew the hydraulic connectors from the pressure and suction
side.
5.
Unscrew the diaphragm rupture sensor from the dosing head.
6.
Remove the 6 screws on the dosing head.
7.
Remove the dosing head.
8.
Check the condition of the diaphragm rupture sensor - see
Ä 'Checking the condition of the diaphragm rupture sensor'
on page 72 .
9.
Loosen the diaphragm from the drive axle with a gentle backwards
turn in an anti-clockwise direction.
10.
Completely unscrew the diaphragm from the drive axle.
11.
Tentatively screw the new diaphragm anticlockwise up to the stop
on the drive axle.
ð The diaphragm is now seated at the stop of the thread while the
diaphragm flap is within the tolerance range.
3
A
Fig. 36: Tolerance range of the flap on the backplate
1
Diaphragm
2
Backplate
3
Flap
A
Tolerance range
Technical personnel
Repair
1
2
P_SI_0029
71