10 Data on electromagnetic compatibility according to EN 60601-1-2 | 10.4 Immunity to electromagnetic interfer-
Rated power P of the
transmitter in W
0.01
0.1
1
10
100
U1 = Compliance level according to 4-6: 3 V
E1 = Compliance level according to 4-3: 3 V/m
Factor
Safe distance depending on the transmission frequency in m
150 kHz to 80 MHz
d=1.17
P
0.12
0.37
1.17
3.69
11.7
eff
[3.5/U1]
For transmitters whose maximum rated power is not included in the above
table, the recommended safe distance d in metres (m) can be calculated using
the equation for the respective column, where P is the maximum rated power
of the transmitter in Watts (W) as specified by the manufacturer.
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
ANMERKUNG 2 Diese Leitlinien mögen nicht in allen Fällen anwendbar sein. The
spread of electromagnetic waves is absorbed and reflected by buildings, objects
and people.

10.4 Immunity to electromagnetic interference

The treatment unitPrimus 1058 Life is designed for operation in an environment
as specified below. The customer or user of the Primus 1058 Life should make
sure that the device is used in an environment of the specified type.
80 MHz to 800 MHz
d=1.20
P
0.12
0.38
1.20
3.79
12
[12/E
]
1
121 / 126
Instructions for use Primus 1058 Life
800 MHz to 2.5 GHz
d=2.3
P
0.23
0.73
2.3
7.27
23
[23/E
]
1