Dräger Fabius Tiro Instructions For Use Manual page 204

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Technical data
Immunity against
Radiated radio fre-
quency disturbance
(IEC 61000-4-3)
Conducted radio fre-
quency disturbance
(IEC 61000-4-6)
1) For P
, insert the highest possible "equivalent isotropic radiated power" of the adjacent radio frequency transmitter. In
EIRP
the vicinity of equipment marked with the symbol
radio frequency transmitters at the location of the medical device should be less than 3 V/m in the frequency range from
150 kHz to 2.5 GHz and less than 1 V/m above 2.5 GHz.
2) ISM bands in this frequency range are: 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to
27.283 MHz; 40.66 MHz to 40.70 MHz.
204
IEC 60601-1-2 test
level
80 MHz to 2.5 GHz:
10 V/m
150 kHz to 80 MHz:
10 V inside ISM
2)
bands
150 kHz to 80 MHz:
3 V outside ISM
2)
bands
, interference can occur. Field strengths from fixed, portable, or mobile
Compliance level
(medical device)
10 V/m
10 V
3 V
Instructions for use Fabius Tiro SW 3.n
Electromagnetic environ-
ment
Recommended minimum
distance to portable and
mobile radio frequency
transmitters with transmis-
sion power P
to the
EIRP
medical device including its
1)
lines:
Recommended minimum
distance from portable and
mobile radio frequency
transmitters with transmis-
sion power P
to the
EIRP
medical device including its
1)
lines: