Cleaning, disinfection and sterilization
Semi-critical medical devices
Components which
can be reprocessed
Breathing system:
– Breathing system
housing
– Inspiratory/expira-
tory ports, APL
valve
– Inspiratory valve,
expiratory valve
– Bag elbow
– Rigid arm for
breathing bag
(optional)
– Flexible arm for
breathing bag
(optional)
– Breathing hoses
Absorber container
and absorber insert
Soda lime dust filter
(optional)
Sample line
– When the sample
line is fitted to the
filter on the Y-
piece.
– When the sample
line is fitted directly
to the Y-piece and
the filters are fit-
ted on the breath-
ing system
Ventilator lid
Ventilator membrane
172
Recommended
Prelimi-
reprocessing
nary
intervals
cleaning
Weekly
Yes
After each patient Observe the associated instructions for use.
Weekly
Yes
Replace each
No
time soda lime is
changed.
Replacement
only
Daily
No
After each patient
After each
Yes
patient
After each
Yes
patient
Machine
cleaning and
Cleaning
disinfection
Yes
Possible
Yes
Possible
No
No
No
No
Yes
No
Yes
No
Instructions for use Fabius Tiro SW 3.n
Manual
Steriliza-
tion
Disinfec-
tion
Possible
Possible
Possible
Possible
No
No
No
No
Yes
Yes
Yes
Yes