Mounting The Cosy Heating And The Power Supply Unit (Optional) - Dräger Fabius Tiro Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Disposable CO
absorber with Drägersorb
2
CLIC (optional)
As an alternative to reusable CO
disposable CO
absorber may also be used.
2
CLIC adapter allows the following single-use CO
absorbers to be used:
– CLIC Absorber 800+
– CLIC Absorber Free
– Infinity ID CLIC Absorber 800+
– Infinity ID CLIC Absorber Free
More information on the connection of Drägersorb
CLIC-Adapters can be found in the associated
instructions for use.
WARNING
Risk of insufficient ventilation
If the CO
absorber is not correctly locked
2
into place, system leakage may occur. The
CO
absorber must audibly engage before
2
Fabius is switched on. This ensures that the
CO
absorber is included into the leakage test
2
and compliance test of the device.
After mounting and replacing, make sure the
CO
absorber is firmly locked into place.
2
Instructions for use Fabius Tiro SW 3.n
absorbers, the
2
2
1 Press the button (G): The CLIC adapter flips
2 Loosen the soda lime in the disposable CO
3 Remove the seal of the new disposable CO
4 Push the disposable CO
Mounting the COSY heating and the
power supply unit (optional)
The compact breathing system can be operated
heated. This reduces the condensation of moisture
in the breathing system during low-flow operation
(fresh-gas flow less than 1 L/min).
 Before installation of the COSY heating, the
Follow the assembly instructions for breathing
system heating (9038262 and 8605899).
Assembly and preparation
open.
absorber, e.g., turn over the disposable CO
absorber a few times.
absorber. Push the disposable CO
the Clic adapter.
upwards into the CLIC adapter, until the CLIC
adapter engages.
leakage test must be performed, see chapter
''Leakage test'' on page 128.
G
H
2
2
2
absorber in
2
absorber (H)
2
65