Ceiling-Mounted Version (Optional) - Dräger Fabius Tiro Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Overview

Ceiling-mounted version (optional)

Fabius can be used in combination with the ceiling
supply units Movita/Movita lift or Forta lift as a
ceiling-mounted version. In this case, the
anesthesia workstation is not placed on a trolley,
(see chapter ''Fabius Tiro as trolley version (front
view)'' on page 20), but rather on the mount of a
ceiling supply unit.
Observe the corresponding assembly instructions
(9037202).
CAUTION
Risk of tipping over
If the device is dismounted from the ceiling
supply unit and set on the floor, it can tip over.
Observe the following steps before setting the
device on the floor:
– Remove all additionally mounted
components.
– Fold the holding arms in on the Fabius.
– Make sure there is sufficient free space.
28
CAUTION
Risk of crushing
When the medical device is set on the floor, body
parts can be crushed.
Make sure there is sufficient free space.
CAUTION
Risk of injury or risk of damage to the medical
device
If the medical device is let down to the floor,
obstructions can damage the medical device. If
the medical device is lifted or lowered, protruding
parts can be damaged.
– Remove all obstructions from under the
device.
– Watch out for protruding parts.
CAUTION
Risk of device failure
If the medical device is used in a tipped position,
parts can be damaged or their function can be
comprised.
Do not use the anesthesia workstation at an
inclination angle over 5°.
CAUTION
Risk of injury or risk of damage to the medical
device
If the maximum weight is exceeded, the holder
can release from the ceiling supply unit.
The total weight of the anesthesia workstation
including all accessories must not be exceeded,
see ''Instructions for mounting the accessories''
and ''Technical data''.