Dräger Fabius Tiro Instructions For Use Manual page 64

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Assembly and preparation
Filling and mounting
D
1 After any cleaning, push the absorber insert
back into the absorber container (D)
completely.
2 Fill CO
absorber with fresh soda lime to the
2
upper mark.
Dräger recommends the use of
Drägersorb 800 Plus or Drägersorb FREE.
WARNING
Risk of hypoventilation
Reuse of the soda lime dust filter can
increase filter resistance and impair the
ventilation function of Fabius.
Replace the soda lime dust filter each time
the soda lime is replaced.
NOTE
Make sure that no soda lime residues are
between the seals and the seal surfaces. This
type of residue can cause system leakages.
64
If conventional, non-pelletized soda lime is used,
then a soda lime dust filter
E
3 Insert a new soda lime dust filter (E). Only use
soda lime dust filters indicated in the list of
accessories.
Only use undamaged filters, as exterior
damage to the filter decreases protection!
4 Place the CO
absorber (F) below the compact
2
breathing system in position and turn
counterclockwise to the final position stop.
CAUTION
Risk of patient recovering consciousness
If the CO
absorber is not correctly locked into
2
place, system leakage may occur.
Make sure that the absorber container is
connected tightly to the compact breathing
system during the ventilation.
*
optional
Instructions for use Fabius Tiro SW 3.n
*
must be used.
F