Dräger Fabius Tiro Instructions For Use Manual page 135

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Restoring the factory settings
Standby Set-up
1 Select Restore Factory Defaults (M) and
confirm.
2 Select Yes or No (N) and confirm.
If Yes is selected, the factory settings are restored.
The factory settings replace the current default
settings.
The values for the factory settings for Fabius are
listed in the following table.
Parameter
Volume Control
Pressure Control
Instructions for use Fabius Tiro SW 3.n
M
N
Factory setting
PMAX= 40
VT= 600
Freq= 12
TI:TE= 1:2.0
TIP:TI= 10
PEEP= 0
PINSP= 15
Freq= 12
TI:TE= 1:2.0
Insp Flow= 30
PEEP= 0
Parameter
Pressure Support
SIMV/PS
Alarm limits for O
2
Alarm limits for MV
Upper alarm limit and
pressure threshold for
PEAK
Minimum Alarm Vol-
ume
Configuration
Factory setting
∆PPS= 10
Freq Min= 3
Trigger= 2
Insp Flow= 30
PEEP= 0
PMAX= 40
VT= 600
Freq= 12
∆PPS= 10
PEEP= 0
Trigger= 2
Insp Flow= 30
TINSP= 1.7
TIP:TI= 10
Upper value = 100
Lower value = 20
Upper value = 12.0
Lower value = 3.0
Upper value = 40
Lower value = 8
Volume = 5
135