Assembly and preparation
Establishing the power supply of the breathing
system heating
1 Plug in the connector of the heating cable (A)
in the corresponding socket (B) on the bottom
of the compact breathing system. When doing
so align the red mark of the plug to the red
mark of the socket.
 After installation of the breathing system
heating, perform the leakage test on Fabius.
If the heating is not used, it must be switched off to
prevent the soda lime from drying out.
WARNING
Risk due to soda lime drying out
The soda lime loses moisture. If the moisture
falls below the minimum moisture, the
following adverse reactions occur
independent of the type of soda lime and
inhalational anesthetic used: Decreased CO
absorption, increased generation of heat in
the CO
absorber resulting in increased
2
breathing gas temperature, formation of CO,
absorption and/or degradation of the
inhalational anesthetics.
– Do not use unnecessarily high fresh-gas
flows.
Only use the supplemental O
necessary.
– Do not leave the flow control valves open
unnecessarily long.
66
B
A
2
delivery if
2
Connecting the compact breathing
system
1 Completely pull out the locking pin (A) and hold
it in that position.
2 Insert the assembled compact breathing
system in the provided holder (B).
3 Release the locking pin again and turn the
compact breathing system until the locking pin
engages.
Instructions for use Fabius Tiro SW 3.n