Setting The Upper Alarm Limit And The Pressure Threshold - Dräger Fabius Tiro Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Setting the upper alarm limit and the
pressure threshold
The standard alarm limits configured for the
ventilation mode can be used unchanged, see
chapter ''Changing the alarm limits'' on page 134.
or
The alarm limits can be set individually for the
current case:
1 Press the
(A) key.
Volume Control
The dialog window (B) with the alarm limits opens.
2 Adjust upper alarm limit and pressure threshold
of the peak pressure (PEAK) (C), see chapter
''Changing the alarm limits'' on page 134.
3 Confirm new values.
Instructions for use Fabius Tiro SW 3.n
A
B
C
C
NOTE
The pressure threshold is preferably to be set so
that the value lies approx. 4 cmH
the current peak pressure.
Monitoring
O (hPa) under
2
123