Stop the frequency modulation.
A large harmonics component is included in the DUT's output signal.
Insert a low-pass filter between the DUT's output terminal and the E5052A RF IN
connector to eliminate the harmonics component.
211
PLL frequency range over
This error occurs when the internal PLL circuit becomes unlocked while the measurement
is in progress. This error may occur when the frequency of the input signal is out of range.
212
PLL Input overflow
This error occurs when the internal PLL circuit becomes unlocked while the measurement
is in progress. This error may occur when the level of the input signal is too large.
200
Power on test failed
This error occurs when the power-on test fails, indicating a failure of the instrument.
Contact an Agilent Technologies sales office or the company from which you bought the
instrument.
40
Printer error
This error occurs when the previous printing is still in progress or the printer fails (offline,
short of paper, etc.) at the time of outputting the display image on the LCD screen to the
printer (HCOP:IMM command).
41
Print failed
This error occurs when printing fails for reasons other than Error 40, Printer error.
-284
Program currently running
This error occurs when the PROG:SEL:STAT RUN command is executed when the VBA
program is in the Run state.
-286
Program runtime error
An error occurring when VBA is executed.
R
75
Recall failed
This error occurs when reading an instrument status file (State01.sta, etc.)
(MMEM:LOAD:STAT command) fails.
220
RF freq out of range
This error occurs when the DUT's output frequency is not within the measurement range.
240
RF level overload
This error occurs when the input to the RF IN port exceeds the maximum input level in the
measurement. The measurement value obtained in such a case is not correct. In the worst
case, a failure (damage to the receiver) may occur.
S
76
Save failed