Table B-9
Status Bit Definitions of Questionable Reference Signal Status Event Register
Bit
Position
0
1
2-15
Issuing the *CLS command will clear all of the bits in the questionable reference signal
status event register.
Table B-10
Status Bit Definitions of Questionable Misc Status Event Register
Bit
Position
0
1
2
3
4-5
6
7-15
Issuing the *CLS command will clear all of the bits in the questionable misc status event
register.
Table B-11
Status Bit Definitions of Questionable Limit Status Event Register
Bit
Position
0
1
2
3
4
5
6-15
Issuing the *CLS command will clear all of the bits in the questionable limit status event
Appendix B
Name
Not used
External Reference Unlock
Not used
Name
Not used
Thermometer Out of Range
IF not Detected
Local Unlock
Not used
AFC Out of Loop
Not used
Name
Not used
Questionable Limit PN1 Status Register
Summary
Questionable Limit SP1 Status Register
Summary
Questionable Limit FP1 Status Register
Summary
Questionable Limit TR1 Status Register
Summary
Questionable Limit USER1 Status
Register Summary
Not used
Status Reporting System
Status Register Structure
Description
Always 0.
Set to "1" when external reference is not locked.
Always 0.
Description
Always 0.
Set to "1" when thermometer is out of range.
Set to "1" when IF is not detected.
Set to "1" when local oscillator is unlocked.
Always 0
Set to "1" when an AFC out of loop error is applied.
Always 0.
Description
Always 0.
Set to "1" when one of the enable bits in questionable
limit PN1 status register is set to "1."
Set to "1" when one of the enable bits in questionable
limit SP1 status register is set to "1."
Set to "1" when one of the enable bits in questionable
limit FP1 status register is set to "1."
Set to "1" when one of the enable bits in questionable
limit TR1 status register is set to "1."
Set to "1" when one of the enable bits in questionable
limit USER1 status register is set to "1."
Always 0.