is not correctly inserted into the drive.
80
File transfer failed
This error occurs when writing data into or reading data from a file (MMEM:DATA
command) fails.
I
213
IF A/D overflow
This message appears when the internal IF level is too high and it exceeds the internal A/D
converter input range. In the phase noise measurement, check that the DUT's carrier signal
is within the selected frequency band. Also, the message may appear if a DUT that has a
large noise level and an unstable output level is measured. When measuring phase noise in
such a case, decrease the IF Gain value or increase the sweep start value if possible. In the
spectrum monitor measurement, decrease the DUT's carrier level, or set the reference level
or input attenuator value larger than the current setting. In the transient measurement,
decrease the DUT's carrier level, or set the max input level or input attenuator value larger
than the current setting.
260
IF Level Overload
The IF level is too high. Set the input attenuator value larger than the current setting.
312
IF not found
The IF signal cannot be found. In the phase noise measurement, this error occurs when no
signal is inputted to the RF IN, or the input level to the RF IN is too low. Also, the message
may appear if a DUT that has an unstable output level is measured. Aside from the above
reasons, there is the possibility of a device failure. Contact an Agilent Technologies sales
office or the company from which you bought the device.
-224
Illegal parameter value
The parameter value is improperly set.
-282
Illegal program name
This error occurs when a nonexistent VBA program name is specified by the
PROG:SEL:NAME command.
-213
Init ignored
Because another measurement is in progress, the request for initiating a measurement
("INIT" command) is ignored.
250
Insufficient IF Level
The IF level is too low. Set the input attenuator value smaller than the current setting.
230
Insufficient RF Level
The input level to the RF IN connector is too low.
L
77
Load VBA program failed
This error occurs when loading a VBA program file fails.
Appendix D
Error Messages
Error number: 80
509