Calculate:sp[1-1]:Trace[1-1]:Hold; Calculate:sp[1-1]:Trace[1-1]:Limit:fail; Calculate:sp[1-1]:Trace[1-1]:Limit:lower:ldata - Agilent Technologies E5052A Programmer's Manual

Signal source analyzer
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

:CALCulate:SP[1-1]:TRACe[1-1]:HOLD

Syntax
:CALCulate:SP[1-1]:TRACe[1-1]:HOLD {OFF|MAXimum|MINimum}
:CALCulate:SP[1-1]:TRACe[1-1]:HOLD?
Description
Selects/Reads trace data hold type
Parameter
OFF(Preset value)
MAXimum
MINimum
Equivalent key
SP Menu -> Trace View -> Data Hold

:CALCulate:SP[1-1]:TRACe[1-1]:LIMit:FAIL

Syntax
:CALCulate:SP[1-1]:TRACe[1-1]:LIMit:FAIL
Description
Reads out the limit test result (Query Only)
Parameter
ON or 1
OFF or 0(Preset
value)
When the limit test is set to OFF, OFF or 0 is always read out.
Equivalent key
No equivalent key is available on the front panel.

:CALCulate:SP[1-1]:TRACe[1-1]:LIMit:LOWer:LDATa

Syntax
:CALCulate:SP[1-1]:TRACe[1-1]:LIMit:LOWer:LDATa <array>
:CALCulate:SP[1-1]:TRACe[1-1]:LIMit:LOWer:LDATa?
Description
Sets/Reads the lower limit values of all measurement points
Parameter
Range
7
:CALCulate:SP[1-1]:TRACe[1-1]:HOLD
Description
Set data hold to 'Off'
Set data hold to 'Maximum'
Set data hold to 'Minimum'
Description
The limit test result is fail
The limit test result is pass
<Description>
1...1024
SCPI Command Reference
161