6.8 Manual calibration procedure
6.8.1 Manual zero calibration
ZERO
CAUTION
For the Ch (components) in which "at once" is
set in the zero calibration (see Section 6.2.2) -
zero calibration is carried out simultaneously.
And for the Ch (components) in which "both"
is set in the calibration range setting
(see Section 6.2.3) - zero calibration is carried
out on both ranges.
Note: For the Ch (component) for
which "AR" is selected in 6.1.1
"Setting of range switch mode",
the cursor automatically moves
to the range selected in "Setting
of auto calibration component/
range" (Section 6.2.4), and cali
bration is carried out within that
range.
To close "Zero Calibration"
To close the "Zero Calibration" or
cancel this mode midway, press the
A previous screen will return.
INZ-TN2ZPB-E
-
-
ENT
-
ENT
ESC
key.
63
Measurement Mode Screen
ZERO
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-50.00
NOx
Range 2 0-500.0
Ch2
Range 1 0-50.00
SO
Range 2 0-500.0
2
Ch3
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-50.00
Ch4
CO
Range 2 0-500.0
Range 1 0-10.00
Ch5
Range 2 0-25.00
O
2
(
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-50.00
NOx
Range 2 0-500.0
Ch2
Range 1 0-50.00
SO
Range 2 0-500.0
2
Ch3
Range 1 0-10.00
CO
Range 2 0-20.00
2
Ch4
Range 1 0-50.00
CO
Range 2 0-500.0
Range 1 0-10.00
Ch5
Range 2 0-25.00
O
2
ENT
ZERO Cal.
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Ch1
Range 1 0-50.00
NOx
Range 2 0-500.0
Ch2
Range 1 0-50.00
SO
Range 2 0-500.0
2
Ch3
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-50.00
Ch4
Range 2 0-500.0
CO
Ch5
Range 1 0-10.00
Range 2 0-25.00
O
2
ENT
To Measurement screen after
executing Manual Zero Calibration
ppm
0.0
ppm
ppm
0.0
ppm
vol%
0.00
vol%
0.0
ppm
ppm
vol%
vol%
20.09
)
ppm
0.0
ppm
ppm
0.0
ppm
vol%
0.00
vol%
0.0
ppm
ppm
vol%
vol%
20.09
ppm
0.0
ppm
ppm
0.9
ppm
vol%
0.34
vol%
ppm
1.1
ppm
vol%
vol%
20.09