6.8 Manual calibration procedure

6.8.1 Manual zero calibration

ZERO
Note
For the Ch (components) in which "at once" is
set in the zero calibration
(see Section 6.2.2) - zero calibration is carried
out simultaneously. And for the Ch
(components) in which "both" is set in the
calibration range setting (see Section 6.2.3) -
zero calibration is carried out on both ranges.
Note: For the Ch (component) for
which "AR" is selected in Sec-
tion 6.1.1 "Setting of range
switch mode", the cursor au-
tomatically moves to the next
range selected in "Setting of auto
calibration component/range"
(Section 6.2.4), and calibration
is carried out within that range.
To close "Zero Calibration"
To close the "Zero Calibration" or
cancel this mode midway, press the
A previous screen will return.
-
-
ENT
-
ENT
ESC
key.
60
Measurement screen
ZERO
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-500.0
NOx
Range 2 0-5000
Range 1 0-500.0
Ch2
Range 2 0-2500
SO
2
Ch3
Range 1 0-200.0
CO
Range 2 0-2000
Ch4
Range 1 0-10.00
O
Range 2 0-25.00
2
(
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-500.0
NOx
Range 2 0-5000
Range 1 0-500.0
Ch2
SO
Range 2 0-2500
2
Ch3
Range 1 0-200.0
CO
Range 2 0-2000
Ch4
Range 1 0-10.00
O
Range 2 0-25.00
2
ENT
ZERO Cal.
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Ch1
Range 1 0-500.0
NOx
Range 2 0-5000
Range 1 0-500.0
Ch2
SO
Range 2 0-2500
2
Ch3
Range 1 0-200.0
CO
Range 2 0-2000
Ch4
Range 1 0-10.00
O
Range 2 0-25.00
2
ENT
To Measurement screen after
executing Manual Zero Calibration
ppm
0.0
ppm
ppm
0.0
ppm
ppm
0.0
ppm
vol%
0.0
vol%
)
ppm
0.0
ppm
ppm
0.0
ppm
ppm
0.0
ppm
vol%
0.0
vol%
ppm
0.0
ppm
ppm
0.9
ppm
ppm
0.1
ppm
vol%
1.1
vol%
INZ-TN2ZPAJ-E

Advertisement

Table of Contents
loading

Table of Contents