Manual Calibration Procedure - Fuji Electric ZPG Instruction Manual

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPG:
Table of Contents

Advertisement

6.8 Manual calibration procedure

6.8.1 manual zero calibration
CAUTION
For the Ch (components) in which "at once" is
set in the zero calibration (see Section 6.2.2) -
zero calibration is carried out simultaneously.
And for the Ch (components) in which "both"
is set in the calibration range setting
(see Section 6.2.3) - zero calibration is carried
out on both ranges.
Note: For the Ch (component) for
which "AR" is selected in Section
6.1.1 "Setting of range switch
mode," the cursor automati-
cally moves to the range selected
in "Setting of auto calibration
component/range" (Section 6.2.4),
and calibration is carried out
within that range.
To close "Zero Calibration"
To close the "Zero Calibration" or
cancel this mode midway, press the
A previous screen will return.
ZERO
ENT
Measurement mode screen
ZERO Cal.
Ch1
NOx
Ch2
O
2
ZERO Cal.
Ch1
NOx
Ch2
O
2
ENT
ZERO Cal.
Ch1
NOx
Ch2
O
2
ESC
key.
To Measurement screen after
executing Manual Zero Calibration
74
ZERO
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Range 1 0-10.00
ppm
Range 2 0-100.0
ppm
Range 1 0-10.00
vol%
Range 2 0-25.00
vol%
(
)
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Range 1 0-10.00
ppm
Range 2 0-100.0
ppm
Range 1 0-10.00
vol%
Range 2 0-25.00
vol%
ENT
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Range 1 0-10.00
ppm
Range 2 0-100.0
ppm
Range 1 0-10.00
vol%
Range 2 0-25.00
vol%
ENT
INZ-TN2ZPG-E
0.0
20.09
0.0
20.09
0.0
20.09