6.7 Manual calibration procedure
6.7.1 Manual zero calibration
ZERO
Note
For the Ch (components) in which "at once" is
set in the zero calibration
(see Section 6.2.2) - zero calibration is carried
out simultaneously. And for the Ch
(components) in which "both" is set in the
calibration range setting (see Section 6.2.3) -
zero calibration is carried out on both ranges.
Note: For the Ch (component) for
which "AR" is selected in Sec-
tion 6.1.1 "Setting range switch
mode," the cursor automatically
moves to the next range selected
in "Setting of auto calibration
component/range" (Section
6.2.4), and calibration is carried
out within that range.
To close "Zero Calibration"
To close the "Zero Calibration" or
cancel this mode midway, press the
A previous screen will return.
-
-
ENT
key, and zero
-
ENT
ESC
key.
50
ZERO
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Ch2
Range 1 0-500.0
Range 2 0-2000
CO
(
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-500.0
Ch2
Range 2 0-2000
CO
ENT
ZERO Cal.
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Ch2
Range 1 0-500.0
CO
Range 2 0-2000
ENT
To Measurement screen after
executing Manual Zero Calibration
vol%
0.00
vol%
0.0
ppm
ppm
)
vol%
0.00
vol%
ppm
0.0
ppm
vol%
0.00
ppm
ppm
0.0
ppm
INZ-TN2ZPAH-E