Manual Calibration Procedure; Manual Zero Calibration - Fuji Electric ZPAH Instruction Manual

Ndir type infrared gas analyzer, replacement for zrh
Table of Contents

Advertisement

6.7 Manual calibration procedure

6.7.1 Manual zero calibration

ZERO
Note
For the Ch (components) in which "at once" is
set in the zero calibration
(see Section 6.2.2) - zero calibration is carried
out simultaneously. And for the Ch
(components) in which "both" is set in the
calibration range setting (see Section 6.2.3) -
zero calibration is carried out on both ranges.
Note: For the Ch (component) for
which "AR" is selected in Sec-
tion 6.1.1 "Setting range switch
mode," the cursor automatically
moves to the next range selected
in "Setting of auto calibration
component/range" (Section
6.2.4), and calibration is carried
out within that range.
To close "Zero Calibration"
To close the "Zero Calibration" or
cancel this mode midway, press the
A previous screen will return.
-
-
ENT
key, and zero
-
ENT
ESC
key.
50
ZERO
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Ch2
Range 1 0-500.0
Range 2 0-2000
CO
(
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Range 1 0-500.0
Ch2
Range 2 0-2000
CO
ENT
ZERO Cal.
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Ch1
Range 1 0-10.00
CO
Range 2 0-20.00
2
Ch2
Range 1 0-500.0
CO
Range 2 0-2000
ENT
To Measurement screen after
executing Manual Zero Calibration
vol%
0.00
vol%
0.0
ppm
ppm
)
vol%
0.00
vol%
ppm
0.0
ppm
vol%
0.00
ppm
ppm
0.0
ppm
INZ-TN2ZPAH-E