11.2.12

Command MI

<id>MI<parameter><CR>
Access:
Description:
Parameter type:
Parameter:
Type return value:
Return value:
Value range:
Initial value:
11.2.13

Command MS

<id>MS<parameter><CR>
Access:
Description:
Parameter type:
Parameter:
Type return value:
Return value :
Value range:
Initial value:
© Adolf Thies GmbH & Co. KG · Hauptstraße 76 · 37083 Göttingen · Germany
Phone +49 551 79001-0 · Fax +49 551 79001-65 · info@thiesclima.com · www.thiesclima.com
Averaging interval
reading / writing
With this command, the averaging interval for floating averaging
of the wind velocity is set in seconds (see Bit4 in the sensor sta-
tus). Wind velocity values are logged 4 times per second (4 Hz)
for averaging.
unsigned integer
e.g. 60  60[s]
unsigned integer
see parameter
0...600
10
Source mean value calculation
read / write
This command is used to set the source for the calculation of
the following values:
- mean value
- gust
- minimum
- standard deviation
- intensity of turbulence
unsigned integer
Parameter
0
Wind velocity calculated by means of standard char-
acteristic curve.
1
Wind velocity calculated by means of calibration table.
When no calibration table is available, the value is
output, which was calculated by means of the stand-
ard characteristic curve.
2
Wind velocity calculated by means of calibration table
and compensated concerning air pressure. When no
calibration table is available, the value is output, which
was calculated by means of the standard characteris-
tic curve.
3
reserved
4
reserved
unsigned integer
see parameter
0...4
0
Description
021814/03/21
Page 43 of 86