11.2.11

Command KY

<id>KY<parameter><CR>
Access:
Description:
Parameter type:
Parameter:
Type return value:
Return value:
Value range:
Initial value:
Note:
After changing the parameter "KX" to a value other than 0, all parameters
(general settings and calibration data) can only be changed by specifying
the password set with "KX"!
Note:
Please contact the manufacturer if you have lost the password ("KX")!
© Adolf Thies GmbH & Co. KG · Hauptstraße 76 · 37083 Göttingen · Germany
Phone +49 551 79001-0 · Fax +49 551 79001-65 · info@thiesclima.com · www.thiesclima.com
Key/password
reading / writing
With this command, the value for the key (password) is set. The
following 3 password levels are possible:
Query (reading)
User (general settings)
Calibration data (access to the calibration data sets)
KX (general settings / access to the calibration data sets)
unsigned integer
Parameter
0
1
234
1
137
Calibration data (see command CK)
2
KX
1)
The key „234" and „137" are valid, if parameter KX the value 0 in-
clude.
2)
Include parameter KX the value unlike 0, then replace the value both
keys „234" (User) and „137" (Calibration data).
unsigned integer
see parameter
0, 137, 234, KX
0
Description
Query
User
User defined password
021814/03/21
Page 42 of 86