Weinmann MEDUMAT Basic Servicing And Repair Instructions page 44

Ventilators
Table of Contents

Advertisement

14. Make the tube connections:
– Push the suction connector
nozzle of the filter holder and onto the con-
nection on the pneumatic block
– Using the ventilation tube
spring
, connect the front nozzle of the fil-
54
ter holder to the pneumatic block
– Make sure that all the tube ends are pushed
on to the limits.
15. Slide the pressure tube
pneumatic block as far as it will go.
16. Slide pressure measurement tube
tube connection of the angled outlet.
17. Secure the circuit board (see „6.12 Replacing
the circuit board" on page 29, steps
).
17.
18. Turn the MEDUMAT Basic / Basic p round.
19. Secure the ventilation control knob
proceed as follows:
– Push the control knob
just short of the limit.
– Twist the knob until the white line is pointing
to the lowest value.
– Hold the knob steady with the special tool
and screw it down.
20. Check the display on the control knob
when turned to the left stop, the white line must
point to the lowest MV value.
If this is not the case, loosen the nuts and align
the control knob
.
35
21.
Perform calibration (see "6.9 Calibration after re-
moval of PCB, ventilation control knob 5 and/or
pneumatic block" on page 25).
22. Close the device (see "6.5 Closing the
device" on page 23).
23. Place the lid
on the knob
34
24. Basic p only:
Carefully screw the cover
switch
.
9
25. Perform a functional check (see "3. Final
Check" on page 8).
44
Repair information and repair instructions
onto the rear
43
.
24
with the
45
.
24
into the inlet of the
48
onto the
50
to
11.
. Please
5
onto the spindle
35
:
35
.
5
onto the two-way
27
48
50
+
43 45
54
Filter holder
35
(Two-way switch only in Basic p)
35
5
34
(Two-way switch only on Basic p)
52

Advertisement

Table of Contents
loading

This manual is also suitable for:

Medumat basic pWm 22600Wm 22650

Table of Contents