Note:
The rocker will tend to drop into the lower
position.
5.
Up to appliance no. 1019 (Basic) or 1399
(Basic p):
Re-install the pneumatic block (see „6.14 Re-
placing the pneumatic block" on page 32,
steps
to
).
10.
20.
From appliance no. 1020 (Basic) or 1400
(Basic p):
Re-install the pneumatic block (see „6.15 Re-
placing the pneumatic block with
angled outlet" on page 34, steps
6. Perform a functional check (see "3. Final
Check" on page 8).

6.18 Replace upper part of housing/control panel

Tools required:
Crosstip screwdriver, size 2,
Crosstip screwdriver, size 1,
L-handled socket wrench 10 mm,
Special tool WM 22829 from special tool set WM 15349,
Special socket spanner 8 mm WM 22826,
Flat nose pliers,
Side nippers,
Cable tie,
Calibration device WM 22836.
Remove upper part of housing
1. Remove the ventilation control knob
proceed as follows:
– Twist the control knob
that you have a reference point when you
come to re-assemble it.
– Lift off the lid
.
34
– Using the special tool, hold the control knob
steady and loosen the nuts with an L-han-
dled socket wrench (10 mm).
– Pull off the control knob
2. Basic p only:
– Unscrew the cap
27
switch
.
9
– Unscrew the nut with the special socket
spanner.
to
).
14.
28.
. Please
5
to the left limit so
35
.
35
from the two-way
35
5
34
(Two-way switch only on Basic p)