6.20 Upgrading ventilation tube

In order to reduce the ventilation resistance of the respiration tube, the ventilation tube
device number 1011 (Basic) or 1137 (Basic p) should be upgraded in the course of the two-yearly service.
In addition a spring
is installed to prevent kinks in the tube.
54
Tools required:
Crosstip screwdriver size 2,
Socket wrench 7 mm,
Side nippers or sharp knife,
Spring
.
54
1. Open the device (see "6.4 Opening the
device" on page 22).
2. Pull the ventilation tube
nection.
3. Unscrew the elbow connection.
4. Unscrew the chromium-plated tube sleeve from
the elbow connection.
5. Screw the chromium-plated tube sleeve togeth-
er with a sealing washer into the connection
(where the elbow connection was previously
installed).
6. Shorten the ventilation tube
65 mm.
7. Slide the spring
onto the ventilation tube
54
8. Push the ventilation tube
9. Close the device (see "6.5 Closing the
device" on page 23).
10. Perform a final check (see "3. Final Check" on
page 8).
46
Repair information and repair instructions
off the elbow con-
45
to exactly
45
.
45
onto the tube nipple.
45
on devices up to
45
Elbow
45
connector
Tube
+
45
54
sleeve