17.
For conversion up to appliance No. 1019 (Basic)
or 1399 (Basic p):
The pressure measuring tube
the housing with a cable tie. Cut through the
cable tie with side nippers and remove the
tube or pull the tube off upwards the housing
with the cable tie.
18. Remove the filter insert:
– Pull the filter cap
55
– Using a screwdriver, press out the pin
– Take the valve insert
cle in the housing e.g. by tilting it with a
small screwdriver then pulling it out with flat
nose pliers.
19. Finally, remove the four rubber buffers
If you are not replacing the control panel, continue
at "Fit upper part of housing".
Replace control panel
1. Remove upper part of housing (see above).
2. Push a screwdriver against the metal panel
from the inside until control panel
loose.
3. Remove adhesive residues completely from the
upper part of the housing.
is secured to
49
out of the housing wall.
.
61
out of the recepta-
57
.
38
comes
65
49
up to appliance no.:
Basic 1019; Basic p 1399
50
from appliance no.:
Basic 1020; Basic p 1400
61
55
57
38
Repair information and repair instructions
21