4.2 Batteries and fuses

MEDUMAT Basic / Basic p is fitted with two bat-
teries which must always be changed together:
A button cell CR2430
17
with auxiliary power if the capacity of the main
battery
is exhausted. This means that an alarm
19
can still be activated in the event of sudden failure
of the main battery. The device switches to expira-
tion.
As a general rule, the capacities of the batteries
are designed in such a way that under normal us-
age conditions, they do not need to be changed
during the 2-year servicing intervals. Within the
context of the prescribed 2-year servicing, the bat-
teries are replaced completely.
We recommend that the batteries be changed
only by the manufacturer Weinmann or by authori-
sed specialists explicity authorised by them, since
special precautions must be taken to protect the
electronics (see "6.6 Changing the batteries" on
page 23)).
For information on replacing the fuse, see Section
„6.7 Replacing the fuse" on page 24.

4.3 Adjusting the pressure gauge

In the idle state, with MEDUMAT Basic / Basic p deactivated and the oxygen cylinder closed, the needle of
the pressure gauge must point precisely to "
To adjust the needle, proceed as follows:
1. Carefully lever out the plastic cover of the ad-
justing screw.
2. Adjust the needle with the adjusting screw
using a small screwdriver.
3. Re-insert the plastic cover.
16
Servicing
supplies the electronics
".
0
:Button cell for auxiliary power
17
:Fuse to prevent internal short-circuiting
18
:Lithium battery 3.6 V for main power to
19
MEDUMAT Basic / Basic p
Adjusting screw
17
18
19