4. Operation

4.1 Automatic switch on/off

Activate automatic mode
M = manual
A = automatic
1.
With the SOMNOcomfort 2e switched off, press the softstart button until the current
setting "M" starts to flash in the display.
2.
Press the on/off switch to change the setting to "A".
If a button is not pressed for 5 seconds, the current setting is saved.
When the automatic system is activated, you can switch SOMNOcomfort 2e on and off
using the on/off switch or using the pressure available in the mask.
SOMNOcomfort 2e is switched on automatically as soon as the patient
breathes into the mask (pressure > 0.5 hPa).
If the mask is removed, the unit will switch off after approx. 15 seconds.
Note:
If the JOYCE Full Face mask is used, the device does not switch on/off automatically,
even if setting "A" is selected. The emergency exhalation valve of the mask means that
the device cannot detect the pressure change required for switching.
Switch off automatic system
M = manual
A = automatic
1.
With the SOMNOcomfort 2e switched off, press the softstart button until the current
setting "A" starts to flash in the display.
2.
Press the on/off switch to change the setting to "M".
If a button is not pressed for 5 seconds, the current setting is saved.
When the automatic system is switched off, you can only switch SOMNOcomfort 2e on
and off using the on/off switch.
If the unit is not used for 15 minutes, it switches off automatically.
Operation
EN
19