8. As soon as you release the Start/Stop button,
communication between the devices is auto-
matically established. While this is happening,
the LEDs 5/16, 17/14 and Start/Stop flash.
Once the Start/Stop LED stay on continuously
and LEDs 5/16 and 17/14 have gone out,
communication is established. The solenoid
valve of the MEDUMAT Basic / Basic p does
not switch any more.
9. Turn the ventilation control knob
MEDUMAT to the setting MV=5,
frequency=16.
10. Press the middle button on the calibration
device. The corresponding LED 5/16 must
light up.
11. Turn the ventilation control knob
MEDUMAT to the setting MV=17,
frequency=14.
12. Press the middle button on the calibration
device. The corresponding LED 17/14 must
light up.
13. Press the bottom button Start/Stop on the cal-
ibration device. All LEDs except Status go out.
You must hear the solenoid valve of the
MEDUMAT Basic / Basic p switching.
14. Switch off the calibration device at the toggle
switch.
15. Disconnect the calibration device from the
MEDUMAT.
16. Close the device (see "6.5 Closing the
device" on page 23).
17. Perform a functional check (see "3. Final
Check" on page 8).
18. Turn MEDUMAT Basic / Basic p off.
26
Repair information and repair instructions
on
5
on
5