Servicing; Intervals - Weinmann homecare SOMNOcomfort 2 Servicing And Repair Instructions

Table of Contents

Advertisement

3. Servicing

3.1 Intervals

Servicing every 2 years or 5000 operating hours
(SOMNOcomfort 2/SOMNObalance only)
Following components to be wiped down with disinfectant:
Unit housing
Filter compartment cover
Cover
Power supply
Power supply cable
Replace filter
Check proper device operation incl. pressure
measuring
Only if excessive soiling is present:
Clean inside using a vacuum cleaner:
Unit housing
Filter compartment cover
Clean areas of extreme soiling.
Servicing every 4 years or 10,000 operating hours / change of patient
Following components to be wiped down with
disinfectant:
Unit housing
Filter compartment cover
Power supply
Power supply cable
Clean inside using a vacuum cleaner:
Unit housing
Filter compartment cover
Clean areas of extreme soiling.
Replace the following parts:
Baffle box, assembled
Cover
5
Gasket, fan housing
Foam, lower part
59
Decoupling tube, outside
Adapter, assembled
Replace filter
Check proper device operation incl. pressure
measuring
10
Servicing
53
66
50
8
(see chapter 4., page 13)
(see chapter 3.2, page 11)
(see chapter 5., page 15)
Opening the unit (see chapter 7.4, page 23)
Closing the unit (see chapter 7.5, page 24)
(see chapter 4., page 13)
Opening the unit (see chapter 7.4, page 23)
Closing the unit (see chapter 7.5, page 24)
(see chapter 7.6, page 25)
(see chapter 3.2, page 11)
(see chapter 5., page 15)