3. Servicing
3.1 Intervals
Servicing every 2 years or 5000 operating hours
(SOMNOcomfort 2/SOMNObalance only)
Following components to be wiped down with disinfectant:
•
Unit housing
•
Filter compartment cover
•
Cover
•
Power supply
•
Power supply cable
Replace filter
Check proper device operation incl. pressure
measuring
Only if excessive soiling is present:
Clean inside using a vacuum cleaner:
•
Unit housing
•
Filter compartment cover
Clean areas of extreme soiling.
Servicing every 4 years or 10,000 operating hours / change of patient
Following components to be wiped down with
disinfectant:
•
Unit housing
•
Filter compartment cover
•
Power supply
•
Power supply cable
Clean inside using a vacuum cleaner:
•
Unit housing
•
Filter compartment cover
Clean areas of extreme soiling.
•
Replace the following parts:
•
Baffle box, assembled
•
Cover
5
•
Gasket, fan housing
•
Foam, lower part
59
•
Decoupling tube, outside
•
Adapter, assembled
Replace filter
Check proper device operation incl. pressure
measuring
10
Servicing
53
66
50
8
–
(see chapter 4., page 13)
–
(see chapter 3.2, page 11)
–
(see chapter 5., page 15)
–
Opening the unit (see chapter 7.4, page 23)
–
Closing the unit (see chapter 7.5, page 24)
–
(see chapter 4., page 13)
–
Opening the unit (see chapter 7.4, page 23)
–
Closing the unit (see chapter 7.5, page 24)
–
(see chapter 7.6, page 25)
–
(see chapter 3.2, page 11)
–
(see chapter 5., page 15)