9. Servicing

9.1 Intervals

We recommend having all ser-
vicing and repair work carried
out by Weinmann or an autho-
rized dealer.
For hygiene reasons we recommend:
Replacing the pressure measurement tube
if dirty;
Replacing the complete mask system every 6 to 12 months depending on
soiling;
Replacing the expiration system in accordance with the relevant operat-
ing instructions.
As a preventive maintenance measure, the device should also be serviced as follows. Either:
after 5000 operating hours (service symbol
or after a maximum of 2 years (see service sticker on back of device).
A service includes:
Changing filters;
Cleaning the device;
Replacing any faulty parts;
Replacing the battery;
Full check on device functions and pressure indicator.
The two filters need to be checked regularly for soiling.
• You should clean the coarse dust filter once a week and
change it at least every 6 months.
• The fine filter must be changed at least every 250 oper-
ating hours.
after 6 months – or earlier
11
appears in the display);
23
Servicing
25