Use the OXYWAY Click distributor only if all the click dials are working
properly.
6. Troubleshooting
Fault
Unusually high oxygen consumption Leak in the system
Inadequate O
supply on inhalation
2
Distributor not working
7. Servicing
7.1 Intervals
Have a control measurement performed every four years to check metering accuracy.
We recommend having repair work performed by the manufacturer, Weinmann, or by
trained experts expressly authorized by Weinmann.
7.2 Method
The device should be cleaned and disinfected beforehand (see "4. Hygiene treatment" on
Page 11).
The following points should be taken into account.
•
Visual inspection:
– mechanical damage
– labeling of components
•
System leaktight
Cause of fault
Leak in the system
Distributor defective or
contaminated from the
inside
Distributor defective
Oxygen source defective or
empty
Remedy
Find and eliminate leak
Find and eliminate leak
Have distributor repaired
Have distributor repaired
Check oxygen source
Troubleshooting
13