Table 6-3: Izs Sequence; Table 6-4: Izs Sequence - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
For each mode there are seven attributes that the MODE can have that control operational details
of the SEQUENCE. They are:
Table 6-3: IZS Sequence Attributes
Attribute No.
1.
2.
3.
4.
5.
6.
7.
Example of enabling sequence #2:
Do a span check ½ hour later every other day, lasting 15 minutes, without calibration.
Table 6-4: IZS Sequence Example
Mode and Attribute
Sequence
Mode
Timer Enable
Starting Date
Starting Time
Delta Days
Delta Time
Duration
Calibrate
Analyzer Instruction Manual, 02246, Rev. F1
X
Attribute Name
Action
Timer Enabled
Turns on the Sequence timer
Starting Date
Sequence will operate after Starting Date
Starting Time
Time of day sequence will run
Delta Days
Number of days to skip between each Seq. execution
Delta Time
Number of hours later each "Delta Days" Seq is to be run
Duration
Number of minutes the sequence operates
Calibrate
Calibrate the instrument at end of sequence
Value
2
4
ON
Sept. 4, 1996
01:00
2
00:30
15.0
NO
Comment
Define Seq. #2
Select Span Mode
Enable the timer
Start after Sept 4
First Span starts at 1:00AM
Do Seq #2 every other day
Do Seq #2 ½ hr later each time
Operate Span valve for 15 min
Do not calibrate at end of Seq
6-4