Maintenance; Maintenance Schedule; Table 8-1: Preventative - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledy ne API Model 200A NO

8 MAINTENANCE

8.1 Maintenance Schedule

The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
Table 8-1: Preventative Maintenance Schedule
Item
TEST Functions
Zero/Span Calibration
Zero/Span Checks
Particulate Filter
Ozone Scrubber on Pump Pack
Filter for Ozone Scrubber
Filter for PermaPure Drier
Zero Air Canister
Reaction Cell Window
Ozone Flow
Sample Flow
Moly Converter
Pneumatic Lines
Factory Calibration
Leak Check
O-rings
Analyzer Instruction Manual, 02246, Rev. F1
X
NOTE
Maintenance Interval
6 - 12 month intervals
Annually or after repairs
Daily
Weekly as needed
Replace charcoal every 3 months Section 4.2.7, Figure 4-2
Replace every 12 months
Replace every 12 months
Refill every 3 months
Clean annually or as necessary
Check every year
Check every year
Check efficiency every 6 months Figure 8-2, Section 8.6
Examine every 12 months, clean
if necessary
Calibrate each year or after
repairs
Check every year
Replace if seal is broken
8-1
Reference Section
Table 9-1
Section 7
Section 7, Table 6-2
Figure 8-1
Figure 4-2
Figure 9-7
Section 8.5
Section 9.3.8, Figure 8-3
Figure 9-9, Section 9.3.7
Figure 9-9, Section 9.3.7
Figure 8-4, Figure 8-5,
Figure 8-6
Section 9.1.6
Section 8.9
Figure 9-9, Figure 9-10