Manual Zero/Span Check Or Cal With Zero/Span Gas In The Sample Port; Table 7-2: Manual Zero - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
7.1 Manual Zero/Span Check or Cal With Zero/Span Gas in
the Sample Port
The calibration of the instrument can be checked or adjusted using gas from the sample port.
This method is often used when the calibration gas is supplied from an external calibrator and
valve system or when the Analyzer does not include the IZS or Z/S valves option.
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero
value.
Table 7-2
port.
Table 7-2: Manual Zero Calibration Procedure - Zero Gas Thru Sample Port
Step Number
1.
2.
3.
4.
5.
Next, enter the expected NO
Analyzer Instruction Manual, 02246, Rev. F1
X
details the zero calibrate procedure with zero gas coming in through the sample
Action
Press CAL
Wait 10 min
Press ZERO
Press ENTR
Press EXIT
and NO span gas concentrations:
x
Comment
The M200A enters the calibrate mode from sample mode. The
zero gas must come in through the sample port.
Wait for reading to stabilize at zero value.
If you change your mind after pressing ZERO, you can still
press EXIT here without zeroing the instrument.
Pressing ENTR actually changes the calculation equations.
M200A returns to sampling. Immediately after calibration,
data is not added to the DAS averages.