Table 9-8: V/F Boards - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Table 9-8: V/F Board Switch Settings
DAC #
0
1
2
3
*Required settings
10 V output or non-isolated current loop
Switch
3
4
5
6
5 V output or isolated current loop
Switch
3
4
5
6
1 V output
Switch
3
4
5
6
100 mV output
Switch
3
4
5
6
Analyzer Instruction Manual, 02246, Rev. F1
X
Sw 1
Sw 2
ON*
OFF*
ON*
OFF*
ON*
OFF*
ON*
OFF*
State
ON
OFF
OFF
OFF
State
OFF
ON
OFF
OFF
State
OFF
OFF
ON
OFF
State
OFF
OFF
OFF
ON
Sw 3
Sw 4
Comment
Use for non-isolated current loop or 10 V output
Use for non-isolated current loop or 10 V output
Use for non-isolated current loop or 10 V output
Use for non-isolated current loop or 10 V output
Comment
Use for isolated current loop or 5 V output
Use for isolated current loop or 5 V output
Use for isolated current loop or 5 V output
Use for isolated current loop or 5 V output
Comment
Use for 1 V output
Use for 1 V output
Use for 1 V output
Use for 1 V output
Comment
Use for 100 mV output
Use for 100 mV output
Use for 100 mV output
Use for 100 mV output
9-44
Sw 5
Sw 6
Sw 7
OFF*
OFF*
OFF*
OFF*

Advertisement

Table of Contents
loading

Table of Contents