Teledyne API Model 200A NO
Table 5-5: DAS Data Channel Editing
Step
Action
1.
Press SETUP-DAS-EDIT
2.
Press PREV/NEXT
3.
Press EDIT
4.
Press SET> (5 times)
5.
Press EDIT
6.
Toggle OFF to ON
7.
Press ENTR
8.
Press EXIT (4 times)

5.3.4 Range Menu

The instrument operates on any full scale range from 50 to 20,000 ppb. The range is the
concentration value that equals the maximum voltage output on the rear panel of the instrument.
If the range you select is between 50 and 2000 ppb the front panel will read the concentration
anywhere from 0 to 2000 ppb regardless of the range selected. If the range selected is from 2001
to 20,000 ppb the front panel will read from 0 to 20,000 ppb. The apparently wider range of front
panel readouts is because the M200A has 2 internal hardware ranges, namely 0-2000 ppb and 0-
20,000 ppb. The analog output is scaled for the range selected, however the front panel reading
can display correct concentrations over the entire physical range.
Only one of the following range choices can be active at any one time.
If the range is changed such that the physical range switches, the
instrument should be recalibrated. Since the gain is 10x lower on the
0-20,000 ppb range, the offsets may be a factor of 10 less also.
There are 3 range modes:
1. Single Range
2. Auto Range
3. Independent Ranges
Analyzer Instruction Manual, 02246, Rev. F1
X
Comment
Enter DAS menu to edit Data Channels
Select Data Channel to edit
Enter the Edit menu for the selected Data Channel
Scroll through Data Channel properties until RS-232
REPORT: OFF is displayed
Edit selected setup property
Change RS-232 REPORT property
Accepts change
Exits back to sample menu
NOTE
5-16