Teledyne API Model 200A NO
All of the default Data Channels sample more than one parameter, for these channels, each
parameter is printed on a separate line.
There is also a compact format. If this attribute is enabled, all 3 concentration parameters are
printed on one line as shown below:
D 94:08:00 0200 CONC : 20.0 120.0 100.0
The parameters are in the order of NO
To change any of the attributes of a particular data channel, the channel attributes are edited
from the front panel. The following table uses the example of Automatic Reporting. Other
attributes can be edited in a similar fashion.

5.5.8 Internal Variables

A list of M200A variables is shown in Table 9-5.
A list of variables and their settings can be requested over the RS-232 port by:
V LIST
The output from this command is long and will not be shown here. The general format of the
output is:
name = value warning_lo warning_hi (data_lo to data_hi)
Where:
name
value
warning_lo
warning_hi
data_lo
data_hi
Analyzer Instruction Manual, 02246, Rev. F1
X
Lists internal variables and values
= name of the variable
= current value of variable
= lower limit warning (displayed if applicable)
= upper limit warning (displayed if applicable)
= lower limit of allowable values
= upper limit of allowable values
, NO
, and NO.
2
x
5-39