Table 7-8: Enabling - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Table 7-8: Enabling CAL-ON-NO2
Step Number
1.
2.
3.
4.
5.
6.
Once CAL-ON-NO2 is enabled the procedure for zeroing and spanning the instrument is
identical to that given in Section 7.2 for Zero/Span Valves.
Analyzer Instruction Manual, 02246, Rev. F1
X
Action
Comment
Press SETUP-
Causes the M200A go to the VARS(Variables) menu
MORE-VARS
Enter the VARS
Password value is 8 1 8
password, then
ENTR
Scroll the VARS
By pressing NEXT or PREV
menu to select the
CAL-ON-NO2
variable
Press EDIT
Once you have entered EDIT, set the value to ON
Press ENTR
ENTR will enable the CAL-ON-NO2 feature
Press EXIT
Each EXIT press returns the M200A to the next higher menu
level. The highest level is the SAMPLE mode
7-10