Table 9-1: Test Functions - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Table 9-1: Test Functions (Continued)
Test Function
AZERO
HVPS
DCPS
RCELL TEMP
BOX TEMP
PMT TEMP
MOLY TEMP
(table continued)
Analyzer Instruction Manual, 02246, Rev. F1
X
Factory Set-Up
Comment
Check value in
AZERO is the current value of the AutoZero circuit reading.
Final Test Values
Values are typically 0 mV -20/+150. For values outside this range,
Table 2-1
check the AutoZero and NO/NO
Diagnostic mode to manually check the AutoZero valve for
correct operation.
450-900 VDC
This represents the scaled-up HVPS programming voltage to the
HVPS. The design of the HVPS precludes taking a single reading
that indicates the health of the supply. Refer to the HVPS
Troubleshooting Section 9.3.8.5 for a procedure for testing the
HVPS. This TEST function is used primarily to set the HVPS
voltage value. A value not in the 450 to 900 volt range indicates
problems with the HVPS supply.
±
2500
200 mV
DCPS is a composite of the +5 and
been arbitrarily set at 2500
of the voltages in the supply is not working. Check the procedures
for diagnosing the Power Supply Module in Section 9.3.5.
±
50
1° C
The reaction cell temperature is controlled to 50° C
computer. It should only read other values when the instrument is
warming up. If the value is outside the acceptable range, go to the
procedure for diagnosing the Reaction cell temp supply in Section
9.3.8.2. The alarm limits are less than 45° C and greater than
55° C.
The Box Temp is read from a thermistor on the Status/Temp board
8-48° C
(01086). It should usually read about 5° C above room temp. The
M200A is designed to operate from 5 to 40° C ambient. Therefore
the box temperature should be in the range of about 10 to 50° C.
Temperatures outside this range will cause premature failures of
components, and poor data quality. Warning limits are < 8° C and
> 48° C.
±
The PMT detector is very temperature sensitive. The PMT
7
1° C
temperature should always be 7° C, except at power-up.
Temperatures more than
problems with the cooler circuit. See Section 9.3.8.4 for PMT
cooler diagnostic and troubleshooting. Warning limits are < 5° C
and > 15° C.
±
Moly temp is controlled by the CPU to 315° C. After cold start it
315
5° C
requires about 30 min to come to temperature. After temperature is
reached temperature should not vary more than
Section 9.3.4 for troubleshooting. Warning limits are < 290° C and
> 320° C.
±
±
1° C from the set point indicate
9-6
valve for cross port leaks. Use
x
±
15 VDC supplies. It has
200 mV. If it is not in this range one
±
1° C by the
±
5° C. See