Table 8-2: Preventative - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledy
ne API Model 200A NO
Table 8-2: Preventative Maintenance Schedule
Item
TEST Functions
Zero/Span
Calibration
Zero/Span Checks
Particulate Filter
Ozone Scrubber on
Pump Pack
Filter for Ozone
Scrubber
Filter for PermaPure
Drier
Zero Air Canister
Reaction Cell
Window
Ozone Flow
Sample Flow
Moly Converter
Pneumatic Lines
Factory Calibration
Leak Check
O-rings
Analyzer Instruction Manual, 02246, Rev. F1
X
Jan
Feb
Mar
Apr
May
Jun
Jul
Aug
Sep
Oct
8-2
Maintenance
Nov
Dec
Interval
6 - 12 month
intervals
Annually or after
repairs
Dail
y
Weekly as needed
Replace charcoal
every 3 months
Replace every 12
months
Replace every 12
months
Refill every 3
months
Clean annually or as
necessary
Check every year
Check every year
Check efficiency
every 6 months
Examine every 12
months, clean if
necessary
Calibrate each year
or after repairs
Check every year
Replace if seal is
broken

Advertisement